RTK/ImageQuality: Difference between revisions

From Openrtk
Jump to navigation Jump to search
No edit summary
Line 21: Line 21:
** Estimate the p and q parameters of Equation 1 in [http://dx.doi.org/10.1007/s003300050710 <nowiki>[Ohnesorge et al, Eur Radiol, 1999]</nowiki>] which comes down to a two-parameters beam-hardening correction and convert it to a lookup table.
** Estimate the p and q parameters of Equation 1 in [http://dx.doi.org/10.1007/s003300050710 <nowiki>[Ohnesorge et al, Eur Radiol, 1999]</nowiki>] which comes down to a two-parameters beam-hardening correction and convert it to a lookup table.
* The algorithm of [http://dx.doi.org/10.1118/1.3477088 <nowiki>[Kyriakou et al, Med Phys, 2010]</nowiki>] may easily be implemented from the existing code in RTK.
* The algorithm of [http://dx.doi.org/10.1118/1.3477088 <nowiki>[Kyriakou et al, Med Phys, 2010]</nowiki>] may easily be implemented from the existing code in RTK.
= Scatter =
* The patient may be approximated by a constant per projection in a first order approximation. This is done using [http://dx.doi.org/10.1118/1.598066 <nowiki>[Boellaard et al, Med Phys, 1997]</nowiki>] in [http://www.openrtk.org/Doxygen/classrtk_1_1BoellaardScatterCorrectionImageFilter.html in rtk::BoellaardScatterCorrectionImageFilter].
* The [http://dx.doi.org/10.1007/s003300050710 <nowiki>[Ohnesorge et al, Eur Radiol, 1999]</nowiki>] algorithm is being implemented in Louvain-La-Neuve (Belgium).
* The adaptive scatter kernel superposition [http://dx.doi.org/10.1088/0031-9155/55/22/007 <nowiki>[Sun and Star-Lack, PMB, 2010]</nowiki>] will be implemented in Louvain-La-Neuve (Belgium).
* Two solutions using a prior scatter-free CT have been investigated by RTK users:
** [http://dx.doi.org/10.1118/1.4923179 <nowiki>[Park et al, Med Phys, 2015]</nowiki>] have implemented the algorithm of [http://dx.doi.org/10.1118/1.3483260 <nowiki>[Niu et al, Med Phys, 2010]</nowiki>].
** Monte Carlo based scatter correction is investigated in Lyon (France).


= Statistical noise =
= Statistical noise =

Revision as of 03:01, 6 August 2015

This page summarizes the existing and the future solutions in RTK for improving image quality of cone-beam (CB) CT images. It is based on discussions at the RTK meeting on image quality.

X-ray source imperfections

  • Geometric blurring can be corrected by the scatter glare correction detailed in the detector imperfections section.
  • Exposure fluctuations from projection to projection are common. They can be corrected by rtk::I0EstimationProjectionFilter which automatically estimates a weighting constant per projection using an histogram analysis. This filter only works if there are pixels in each projections that measure x-rays that traversed air only (except maybe a few projections using a revursive least-square (RLS) algorithm). The filter does not have any parameter except the bitshift template value for the reduction of the histogram size. It is implemented for integer pixel types only.
  • Focal spot motion cannot be corrected currently. It would require geometric calibration for each acquired projection using auto calibration.

Detector imperfections

  • Variations of the flat field image and the dark field image is known (due to temperature changes or ghosting, see, e.g., [Siewerdsen and Jaffray, Med Phys, 1999]). Acquisition of these two images before and after the acquisition is the best solution when it is possible. There is no other solution in RTK except for the automatic detection of the constant I0 value (see fluctations of the source exposure).
  • Lag corresponds to a short term effect of the detector. The thesis of Starman (2010) gives a good overview of the problem and solutions he proposed. | rtk::LagCorrectionImageFilter implements equation 2.1 of his PhD thesis. The a and b parameters must be calibrated for a given system, some RTK users have done this and could share there scripts upon request via the RTK user mailing list. A CUDA version of this filter is being implemented in Louvain-La-Neuve (Belgium). Solution in chapter 3 of Starman's thesis might also be investigated in future works.
  • Scatter glare is the point spread function (PSF) of the detector. The solution described in [Poludniowski et al, PMB, 2011] is implemetend in rtk::ScatterGlareCorrectionImageFilter. The a and b parameters must be calibrated for a given system, some RTK users have done this and could share there scripts upon request via the RTK user mailing list.

Beam hardening

Scatter

Statistical noise

  • RTK has a fast 2D median filter for projection images for a few kernel dimensiosn, see rtk::MedianImageFilter. A GPU version of the median filter will be developed in Salzburg (Austria).
  • Median filters do not preserve edges (see [Arias-Castro and Donoho, Annals of Statistics, 2009]. A multi-pass median filter is required which might be investigated in Louvain-La-Neuve (Belgium) in the future.
  • The Savitzky–Golay filter is a promising solution that will be investigated in Louvain-La-Neuve (Belgium) in the future. This solution also provides derivatives of the image.

Truncated projection images