This page summarizes the existing and the future solutions in RTK for improving image quality of cone-beam (CB) CT images.

X-ray source imperfections

  • Geometric blurring can be corrected by the scatter glare correction detailed in the detector imperfections section.
  • Exposure fluctuations from projection to projection are common. They can be corrected by rtk::I0EstimationProjectionFilter which automatically estimates a weighting constant per projection using an histogram analysis. This filter only works if there are pixels in each projections that measure x-rays that traversed air only (except maybe a few projections using a revursive least-square . The filter does not have any parameter except the bitshift template value for the reduction of the histogram size. It is implemented for integer pixel types only.
  • Focal spot motion cannot be corrected currently. It would require geometric calibration for each acquisition.

Detector imperfections

  • Lag: the [ rtk::] filter implements correction implements

, see \[[Poludniowski et al, PMB, 2011\] implemetend in [rtk::]

Beam hardening

Truncated projection images